• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BELMONT MEDICAL TECHNOLOGIES THE BELMONT RAPID INFUSER, RI-2; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BELMONT MEDICAL TECHNOLOGIES THE BELMONT RAPID INFUSER, RI-2; PUMP, INFUSION Back to Search Results
Catalog Number 903-00006P
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
The risk manager reported that while using the belmont, the tubing was reported as broken while infusing blood products and "spilling" out of the top of the canister.There is no report of any patient harm associated with this incident.
 
Additional Manufacturer Narrative
The user facility did not inform belmont about this incident when it occurred on (b)(6) 2023, belmont became aware of this incident after receiving medwatch report #: (b)(6) on december 19th, 2023.Internal complaint file #: (b)(4) has been logged for this incident for traceability.The ri-2 and disposable involved in this incident were not returned to belmont for evaluation.The user facility claimed that there was a leak however the disposable in question has not been provided for review nor has the lot number of the device been provided.A leak would be obvious to the user while interacting with the device.The disposable set can be exchanged to continue infusion.There is no report of any patient harm associated with this incident.The operator's manual provides instructions on installing the disposable set and additional recommended operator actions.The step-by-step procedures in manual has warning on installing the disposable, "do not kink or twist the tubing" and "do not apply excessive pressure to the pressure transducer.The pressure transducer can be damaged with excessive force.Do not use the system if the pressure transducer is damaged".All 3 spike sets are 100% visually inspected and 100% leak tested prior to final packaging and release for shipment from belmont medical technologies.Belmont has reached out to the risk manager (initial reporter from the user facility) on december 19th, 2023, january 4th, 2024 and january 12th, 2024 to request the lot number and additional information about the incident but haven't received any further updates on the incident.There is no information about the lot number of the device to allow the review of device manufacturing records.No device malfunction could be verified, and the root cause could not be determined.To date we have not received any response from the user facility.The complaint file will be re-opened in the event the device and additional information is received.Belmont will continue to monitor this issue moving forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE BELMONT RAPID INFUSER, RI-2
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer (Section G)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
Manufacturer Contact
urja jani
780 boston road
billerica, MA 01821
MDR Report Key18515272
Report Number1219702-2024-00004
Device Sequence Number18085892
Product Code FRN
UDI-Device Identifier00896128002022
UDI-Public00896128002022
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K141654
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number903-00006P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2023
Initial Report FDA Received Date01/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-