|
The device was returned and evaluated by olympus.In addition to the findings in b5, evaluation found the knife was not facing the 11 o'clock direction when inserted into a scope.The knife wire was not deformed and there was no abnormality in the length of the wire or sheathing.The tip of the tube was deformed and pointed to the left.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause of the torn coating could not be determined, likely factors causing the tear could have been the following: - it is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.Although a definitive root cause of the wire not pointing in the 11 o'clock direction, it is possible that a force was applied to the distal end of the product during device handling as described below.This may have caused the cutting wire to deform.- when pre-curved stylet was removed - when an attempt was made to curve the distal end of the tube - when the device was inserted into the endoscope - when an attempt was made to extend the distal end of the tube while the forceps elevator was raised - when the guide wire was inserted into the distal end of the tube, if the guide wire was used for the procedure the customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : - do not use excessive force to bend the distal end of the sphincterotome.This could damage the cutting wire.- do not insert the instrument into the endoscope when the cutting wire is tightened.Doing so may extend the distal end of the insertion portion from the endoscope tip abruptly.This could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.- when inserting the instrument into the endoscope, hold the slider firmly.Otherwise, the distal end of the insertion portion may bend and could extend from the distal end of the endoscope abruptly.This could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope or instrument.- do not force the instrument if resistance to insertion is encountered.Reduce the angulation or lower the forceps elevator of the endoscope until the instrument passes smoothly.The use of excessive force could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.- do not advance or extend the instrument abruptly.This could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.- when inserting the instrument into the endoscope, hold it close to the biopsy valve and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.- insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.- when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.- do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|