Model Number N/A
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Joint Dislocation (2374); Insufficient Information (4580)
Date of Event 10/27/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent a partial knee arthroplasty and subsequently a revision surgery was performed due to unknown reasons.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
Cmp-(b)(4) this follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10.No product was returned or pictures provided, so visual and dimensional evaluations could not be performed.It was reported that the patient experienced dislocation of the bearing following a fall.The surgeon performed revision and upsized the bearing 1 mm and was happy with the tension.No other oxford implants were removed or revised.The surgeon reported no user error and that the dislocation was not due to product failure.The surgeon believes that the patient fell, potentially did some damage to her mcl and thus achieved better tension with a poly that was 1 mm thicker than the previous one.Therefore, it is determined that the product meets the applicable acceptance criteria and is conforming to the specifications.Device is used for treatment.The root cause of the reported issue is unrelated to the zimmer biomet medical device.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that patient underwent a partial knee arthroplasty and subsequently a revision surgery was performed due to bearing dislocation.The surgeon believes that the patient fell and reported no surgeon error or implant failure.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG RT MD SIZE 6 PMA
Common Device Name
OXFORD HXLPE BEARINGS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18516064
Report Number3002806535-2024-00022
Device Sequence Number6751254
Product Code NRA
UDI-Device Identifier05019279786244
UDI-Public(01)05019279786244(17)241216(10)056640
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/16/2024
Device Model NumberN/A
Device Catalogue Number159578
Device Lot Number056640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2023
Supplement Date Received by Manufacturer02/08/2024
03/01/2024
Initial Report FDA Received Date01/16/2024
Supplement Report FDA Received Date02/19/2024
03/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-