MEDOS INTERNATIONAL SARL HD LAPSCP,EP,5.5,0,300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
242632 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the healthcare professional that during testing on (b)(6) 2023 it was observed that two scopes were cloudy and were unable to be used.There was no patient involvement reported.No surgical delay or patient consequences reported.No additional information was provided.This is report 1 of 2 for the same event in (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: udi: (b)(4).D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary the complaint device was received at the manufacturing site and evaluated.During the service evaluation the following defects were identified: outer tube damaged,outer tube bent,distal tip damaged.Broken lenses in optical system.Per service reports, this complaint can be confirmed.The defective parts needs to be replaced to resolve the issues.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review a manufacturing record evaluation was performed for the finished device serial number, and no non-conformances were identified.
|
| |
|