MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY ® 2; CATHETER, INTRAVASCULAR, THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY ® 2; CATHETER, INTRAVASCULAR, THE Back to Search Results
Catalog Number 4232006-02
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: brief inquiry description defective iv catheters detailed inquiry description clinicians noted on the iv catheters that there is a "coating" on it.Was described as "waxy" or a "film" on the catheter.Others described as a "ball" like substance appearing on the catheters.Some noted the coating before placing the iv.Others were noted when placed in the patients.Some clinicians stated that when the catheter was threaded into the patient, the coating gathered at the site of insertion.Description of the patient event when threading the iv catheter, the clinician noted the film substance gathering at the insertion site of the catheter.As if the film was coming off the catheter when it was inserted and was gathering at the insertion site.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One sample was submitted to the manufacturer for evaluation.Through visual examination under 20x magnification using a keyence scope on the capillary surface of the received sample, no abnormalities were observed.No foreign material was observed.A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.This product is assembled on an automated assembly machine equipped with vision systems and test stations.Silicone oil is applied on the external capillary surface during siliconization process, and this is to facilitate smooth capillary insertion process.The process cards of the complaint batch show no abnormalities.Based on the investigation, the sample is within specification and the defect could not be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Additional Manufacturer Narrative
Additional information/evaluation: this report has been identified as b.Braun medical internal report number (b)(6).One sample was submitted to the manufacturer for evaluation.Through visual examination under 20x magnification using a keyence scope on the capillary surface fo the received sample, no abnormalities were observed.No foreign material was observed.Two photos were submitted to the manufacturer for evaluation.Through visual examination of the photos, it was observed that the photos are of the product introcan safety 2 wl pur m 20gx25mm - us (catalog # 4242006-02).Due to this, it was determined that the photos do not correspond with the reported products from complaints (b)(4).A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.This product is assembled on an automated assembly machine equipped with vision systems and test stations.Silicone oil is applied on the external capillary surface during siliconization process, and this is to facilitate smooth capillary insertion process.The process cards of the complaint batch show no abnormalities.Based on the investigation, the samples are within specification and the defect could not be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY ® 2
Common Device Name
CATHETER, INTRAVASCULAR, THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key18518422
Report Number9610825-2024-00004
Device Sequence Number6751319
Product Code FOZ
UDI-Device Identifier04046955341894
UDI-Public(01)04046955341894
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K192676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4232006-02
Device Lot Number23A04G8274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2024
Supplement Date Received by Manufacturer01/03/2024
01/03/2024
Initial Report FDA Received Date01/16/2024
Supplement Report FDA Received Date02/02/2024
04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-