| Model Number |
100/897/080 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Material Too Soft/Flexible (4007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).G5: 510k number is unknown.H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the white end of the device had become loose and was not secure.The entrance to the tracheostomy appeared very soft.The device was in use on the patient at the time of discovering the issue.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: one sample was received without its original packaging.A visual inspection found the sample in good condition but the connector was easily disconnected from the tube.The customer reported issue was confirmed.Root cause of this issue remains unknown.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|