• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX TUBES PDT UNIPERC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX TUBES PDT UNIPERC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 100/897/080
Medical Device Problem Codes Loose or Intermittent Connection (1371); Material Too Soft/Flexible (4007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).G5: 510k number is unknown.H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the white end of the device had become loose and was not secure.The entrance to the tracheostomy appeared very soft.The device was in use on the patient at the time of discovering the issue.There was no patient harm.
 
Additional Manufacturer Narrative
Device evaluation: one sample was received without its original packaging.A visual inspection found the sample in good condition but the connector was easily disconnected from the tube.The customer reported issue was confirmed.Root cause of this issue remains unknown.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES PDT UNIPERC
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL CZECH REPUBLIC A. S
olomoucka 306
hranice 753 0 1
EZ  753 01
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18518570
Report Number3011237704-2024-00016
Device Sequence Number12687284
Product Code BTO
UDI-Device Identifier15019315107207
UDI-Public(01)15019315107207(17)280725(10)4382205
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
UNKNOWN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/897/080
Device Catalogue Number100/897/080CZ
Device Lot Number4382205
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date01/16/2024
Supplement Report FDA Received Date08/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-