| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a fractured provisional was discovered during kit inspection.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 foreign source: canada
customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Proposed component code is mechanical (g04) - provisional bottom.Complaint sample was evaluated and the reported event was confirmed.Review of returned instrument found it to exhibit signs of repeated use (nicked / gouged) and a corner of the post feature has fractured off.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|