| Catalog Number |
EPH8716H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When did the needles detach/pull off from the suture (in the package, during removal from the package, during handling before use on patient or during use on the patient)? please specify please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep).Events reported via: mwr-29122023-0001543085, mwr-29122023-0001543086.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the suture was detached from the needle easily.It was not a control release needle.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/9/2024.H6 component code: g07002 no device problem found.Additional information was requested, the following was obtained: when did the needles detach/pull off from the suture (in the package, during removal from the package, during handling before use on patient or during use on the patient)? please specify: unk.Please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep): (b)(6).Please perform and document the follow-up attempt for product return.Please document the shipment tracking number in the notes or rmao sections: the device has been received at sukagawa and will be shipped.Please check rmao.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample determined that it was received, one open sample that pertained to the product code eph8716h.Upon visual inspection of the returned sample, it was noted that the suture was cut in two sections.The swage and attachment were noted to be as expected.The suture pieces were examined, and the extremes were noted to be cut probably caused by a surgical instrument.Also, the functional test was performed using instron equipment and the pull force result was above the minimum requirements.This report is not intended to deny that you experienced a problem with the device.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|