• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 4007
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during infusion, while immobilizing with an iv3000 1 hand 6x7cm ctn 100 dressing, the package was discovered to be broken and was immediately replaced with no harm to the patient.Upon feedback it was informed that this may have been caused by storage problem by the hospital.No further information is available at this time.
 
Additional Manufacturer Narrative
Internal complaint number: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.There are strict inspections during the manufacturing process to ensure that products that do not meet specifications are rejected.Based on feedback the probable cause is hospital storage problem.A review of the batch manufacturing records could not be performed as the lot number provided (20211001) is not a valid lot number for this product.However, there are no indications to suggest that the device did not meet specifications upon release to distribution.A review of complaint history did not reveal similar events for this nature with no manufacturing problems observed.A risk management review concluded that the alleged failure mode and any associated harm has been anticipated within the risk file, with no updates required.A review of prior escalation actions found no actions applicable to the scope of this case.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 6X7CM CTN 100
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18521781
Report Number8043484-2024-00004
Device Sequence Number6731936
Product Code KGX
UDI-Device Identifier05000223417383
UDI-Public05000223417383
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4007
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2023
Supplement Date Received by Manufacturer02/21/2024
Initial Report FDA Received Date01/16/2024
Supplement Report FDA Received Date02/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-