| Catalog Number |
1000186 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in an unspecified artery.The indeflator was used to inflate a non-compliant (nc) balloon at 12 atmospheres (atms) when it started to lose pressure and then the gauge broke apart separating from the housing.Another indeflator was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported material separation was able to be confirmed.The reported leak was unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, it is possible that inadvertent mishandling during preparation for use and/or during use resulted in the noted multiple scratches throughout the indeflator housing; thus compromising the indeflator such that during use resulted in the reported leak and the reported material separation /noted faceplate/gauge separations and the noted housing separation; however, this cannot be confirmed.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Search Alerts/Recalls
|