• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INFINITI VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE INFINITI VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750852
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that loose tip was observed.Additional information requested and received that the event was occurred during the cataract [with intraocular lens (iol) implant] surgery.The surgery was completed on same day.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.10.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM, HANDPIECE TIP
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18523907
Report Number2523835-2024-00036
Device Sequence Number7802007
Product Code HQC
UDI-Device Identifier00380657508525
UDI-Public00380657508525
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K021566
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065750852
Device Lot NumberASKU
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/21/2023
Supplement Date Received by Manufacturer03/24/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK SURGICAL PROCEDURE PACK.; INFINITI VISION SYSTEM, ACCESSORY, HANDPIECE.
Patient SexUnknown
-
-