Model Number VTICM5_13.2
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2, -10.5/3.0/090 (sphere/cylinder/axis), implantable collamer lens into the patient's left eye (os).Excessive vault was observed.Lens remains implanted.Additional information has been requested but none has been forthcoming.
 
Additional Manufacturer Narrative
H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
 
Additional Manufacturer Narrative
Lens was explanted and received.Additional information has been requested but none has been forthcoming.Claim # (b)(4).
 
Additional Manufacturer Narrative
H3 - device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage.Dimensional inspection found the lens to be within specifications.Claim #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key18525777
Report Number2023826-2024-00173
Device Sequence Number20435930
Product Code QCB
UDI-Device Identifier00841542116688
UDI-Public00841542116688
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2023
Supplement Date Received by Manufacturer06/25/2024
07/15/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date06/25/2024
07/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-