| Catalog Number |
121181-000070 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the connector got detached from the tube by itself during use on a patient, immediately after insertion.Therefore, the user fixed it with zip tie.No injury to the patient occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned to the manufacturer.The manufacturer reported: "an actual sample was received for further evaluation.Based on evaluation, the tube can be detached from the connector.The tube was observed with no glue marking on inner side of tube and outer side of connector.Further review on product catalog, the assembled tube was supplied by external supplier.Based on the investigation, the defect mode matches with the complainant.Hence, further investigation and corrective action will be documented." teleflex will continue to monitor and trend complaints of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that the connector got detached from the tube by itself during use on a patient, immediately after insertion.Therefore, the user fixed it with zip tie.No injury to the patient occurred.
|
| |
|