(b)(4).Date sent: 1/17/2024.D4: batch # 197a62.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce60a device was returned with no apparent damage and with two gst60b reloads(b and c) present.Both reloads were received fully fired.In addition, the reload(b) was noted to be bent, with the pan detached from the reload.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb board was noted to be non-functional.Regarding the condition of the reload(b), it is also possible that handling by the customer could dislodge the pan and damaged cartridge.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.Please reference the instruction for use for more information.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the pcb board to be damaged.A manufacturing record evaluation was performed for the finished device batch number 197a62, and no non-conformances were identified.
|