ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2742-9
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A user facility clinic manager (cm) reported to fresenius that the injection port on the saline line of the combiset bloodlines leaked during a patient's hemodialysis (hd) treatment resulting in air in the circuit.The patient's blood was not returned.Additional information was obtained during follow-up with the cm and a user facility registered nurse (rn).The reported issue was detected approximately 1h45 minutes after treatment initiation.The saline line was found to be dripping.The 2008t machine alarmed with an air detector message.The rn stated the entire circuit had filled with air.The patient¿s blood was not returned.The patient¿s estimated blood loss (ebl) was approximately 300 ml.The patient did not experience a serious injury or require medical intervention as a result of the reported issue.The patient completed treatment on the same machine with new supplies.The bloodline set was discarded and is not available to be returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.However, a photograph of the alleged malfunction was provided by the customer.In the photograph, it can be observed that the ¿y¿ injection site of the priming line is leaking.The reported issue is confirmed.The priming line used in the arterial line is a component that is manufactured by an external supplier; therefore, this kind of failure mode could only be caused by the external supplier.A notification was made to the external supplier about the alleged incident related to the injection site.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.
 
Event or Problem Description
A user facility clinic manager (cm) reported to fresenius that the injection port on the saline line of the combiset bloodlines leaked during a patient's hemodialysis (hd) treatment resulting in air in the circuit.The patient's blood was not returned.Additional information was obtained during follow-up with the cm and a user facility registered nurse (rn).The reported issue was detected approximately 1h45 minutes after treatment initiation.The saline line was found to be dripping.The 2008t machine alarmed with an air detector message.The rn stated the entire circuit had filled with air.The patient¿s blood was not returned.The patient¿s estimated blood loss (ebl) was approximately 300 ml.The patient did not experience a serious injury or require medical intervention as a result of the reported issue.The patient completed treatment on the same machine with new supplies.The bloodline set was discarded and is not available to be returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key18527138
Report Number0008030665-2024-00058
Device Sequence Number6469492
Product Code FJK
UDI-Device Identifier00840861100309
UDI-Public00840861100309
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K962081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03-2742-9
Device Lot Number23NR01191
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer02/22/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008T MACHINE.; FRESENIUS 2008T MACHINE.; FRESENIUS BLOODLINES.; FRESENIUS BLOODLINES.
Patient SexFemale
-
-