| Catalog Number |
MX7827 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: month and year of event have been provided, day is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing set off the high-pressure alarm in the pump.The alarm could not be cleared no matter how many times they increased the pressure limits.There was patient involvement, but no patient harm/adverse event reported.The outcome of the event was reported as ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
D9: date returned to mfg.: 1/22/2024.H3 and h6.Evaluation codes: updated.Two samples used outside its original package were received for investigation.The complaint was not confirmed for the failure mode reported ¿tubing set off the high-pressure alarm in the alaris large volume pump that couldn't be cleared¿.Through the device analysis executed it was found that the returned sample passed the electrical test.Awareness for failure ¿tubing set off the high-pressure alarm in the alaris large volume pump that couldn't be cleared¿ was given to personnel who perform the assembly process.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.
|
| |
|