• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MASTERS SERIES HEART COATED AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MASTERS SERIES HEART COATED AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 25CAVGJ-514 00
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 12/26/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023, a 25mm sjm masters series coated aortic valved graft was selected for use in a bentall procedure.During the procedure, the patient's activated clotting time (act) level was 100-110 seconds.A blood transfusion was required.It was noted during procedure that after the valved graft was implanted, there was blood leakage throughout the area, including the area where the artificial valve and artificial blood vessel of the valved graft were connected.The area was reinforced with a pericardial patch and hemostatic material.After the procedure, the patient's international normalized ratio (inr) was 1.8.The physician was not sure if the cause of the bleeding was due to a leak from the blood vessel or a gap between the blood vessel and the valve.The patient was reported as stable and was later discharged.
 
Additional Manufacturer Narrative
An event of blood leakage throughout the area, including the area where the artificial valve was reported.It was also reported that the area was reinforced with a pericardial patch and hemostatic material.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES HEART COATED AORTIC VALVED GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18527622
Report Number2135147-2024-00273
Device Sequence Number6469001
Product Code LWQ
UDI-Device Identifier05414734006316
UDI-Public05414734006316
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number25CAVGJ-514 00
Device Lot Number9190460
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/27/2023
Supplement Date Received by Manufacturer01/22/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexMale
-
-