U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063901010
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of basket wire detached.
 
Event or Problem Description
It was reported to boston scientific corporation that a zero tip basket device was used during a ureteroscopic lithotripsy (ursl) procedure performed on (b)(6) 2023.During the procedure, basket wire and the handle completely detached.The pieces were removed upon separation.The procedure was completed with another zero tip basket device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of basket wire detached.Block h10: investigation results the returned zero tip basket was analyzed, and a visual evaluation noted that the cannula and sheath returned detached from the handle.It was also noted that the basket was slightly extended and there is no evidence of basket wires broken.With all available information, boston scientific concludes that the reported event of basket wire break is not confirmed the since condition of the device does not let to perform a functional inspection.Based on the investigation results this could be related to handling and manipulation of the device during procedure, it is probable that an excess of force applied to the device such as operational factors during their use could lead to cause the detachment of the cannula and sheath from the handle.Also, instruction for use (ifu) emphasizes that if resistance is encountered during advancement or withdrawal of the device, do not exert excessive force.Based on the investigation results, the most probable cause of adverse event related to procedure was assigned to this investigation.
 
Event or Problem Description
It was reported to boston scientific corporation that a zero tip basket device was used during a ureteroscopic lithotripsy (ursl) procedure performed on (b)(6) 2023.During the procedure, basket wire and the handle completely detached.The pieces were removed upon separation.The procedure was completed with another zero tip basket device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZERO TIP
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18528300
Report Number2124215-2024-01716
Device Sequence Number6383151
Product Code FFL
UDI-Device Identifier08714729342007
UDI-Public08714729342007
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2025
Device Model NumberM0063901010
Device Catalogue Number390-101
Device Lot Number0030703090
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/28/2023
Supplement Date Received by Manufacturer02/09/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date03/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-