| Model Number |
V+O+C+S+N+PRO, JAPANESE |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to ventec that the ventilator was unable to maintain peep (positive end expiratory pressure).There were no reports of patient involvement associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: ventec will perform an evaluation of the device.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was not returned to ventec for evaluation.The device was evaluated by an authorized service provider (asp) where the reported issue of it being unable to maintain peep (positive end expiratory pressure) was confirmed.The asp replaced the internal flow transducer (ift) to resolve the reported issue.Proper device operation was then observed through functional and performance testing.The investigation determined that the cause of the reported issue was the ift.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information for supplemental medwatch report 002 -- section d4, model #, of the initial medwatch report stated: prt-00490-100.Section d4, model #, of the initial medwatch report should have stated: v+o+c+s+n+pro, japanese.Section d4, udi #, of the initial medwatch report stated: (b)(4).Section d4, udi #, of the initial medwatch report should have stated: (b)(4).
|
| |
|