AVANOS MEDICAL INC. MICROCUFF ADULT ENDOTRACHEAL TUBE; VAP MICROCUFF ENDOTRACHEAL TUBE Back to Search Results
Model Number 35216
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
4581-appropriate clinical signs, symptoms, conditions term/code not available: patient coded.The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 16 jan 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event or Problem Description
Fda user facility med watch report fda mw (b)(4) reported, "the 15mm connector, where the ventilator or o2 [oxygen] source connects, disconnects from the neck of the ett [endotracheal tube].The 15mm connector appears to become loose from the neck of the ett when warmed in the airway.When this occurs, the patient is not effectively being ventilated.At least one patient has had to be reintubated in order to continue ventilator treatment.Overnight on [redacted date], there was a patient in the medical icu [intensive care unit] who coded post intubation.While taking the colormetric off and placing the emma [emergency capnometer], the blue piece just popped off.The respiratory therapist was able to get the connector back on quickly." it was additionally reported, the event occurred in the patient¿s hospital room and the device had been in place for approximately 1 day; the patient's current condition was not reported.
 
Additional Manufacturer Narrative
The device history record for lot cm3118001 was reviewed and the product was produced according to product specifications.All information reasonably known as of 04 mar 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF ADULT ENDOTRACHEAL TUBE
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBE
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
COMMAND MEDICAL NICARAGUA SA
km 12.5 carretera norte
parque industrial las mercedes, edificio 16
managua, managua
Manufacturer Contact
ujjal chakravartty
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key18530546
Report Number3011270181-2024-00011
Device Sequence Number15378125
Product Code BTR
UDI-Device Identifier00609038352169
UDI-Public00609038352169
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K113333
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number35216
Device Lot NumberCM3118001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/26/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-