| Catalog Number |
0502937030 |
| Medical Device Problem Codes |
Fogging (1253); Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was previously closed based on product not received; however, the product has now been physically received at stryker endoscopy, usa and the investigation has been re-opened.Investigation as follows is now based on product received.Alleged failure: it has been reported that the scopes does not transmit a clear imagen, when they are sterilized appear as drops of steam inside the scopes.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be rough handling during sterilization/product transport, or fluid ingress.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image during procedure.
|
| |
|
Search Alerts/Recalls
|