Catalog Number 941-01-40H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to infection.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
See d4expiration date.The reason for this revision surgery, the agent reported "(patient presented with septic symptoms just 2 weeks after primary joint was implanted)".The previous surgery and the surgery detailed in this event occurred 13 days apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported device was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was an nonconformance associated with the main part #941-01-40h, empowr acetabular system, liner, neutral, hxe+, 40h which documents that out of 40 parts lot 1 part was rejected and scrapped.All other items in the lot were met with fit, form and function requirements.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to septic symptoms.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions or incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR ACETABULAR SYSTEM, LINER, NEUTRAL, HXE+, 40H
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key18531773
Report Number1644408-2023-02024
Device Sequence Number12457296
Product Code LPH
UDI-Device Identifier00190446309598
UDI-Public00190446309598
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K190057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number941-01-40H
Device Lot Number605Z1198
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2023
Supplement Date Received by Manufacturer01/18/2024
Initial Report FDA Received Date01/17/2024
Supplement Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-402 LOT: 871B1454
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-