| Model Number |
42-702 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The inspection of further torque wrench devices by the supplier showed a sporadic untight welding seam, which allowed liquids (e.G.Cleaning or disinfection agents) and/or steam to enter the inside of the torque wrench during processing.Thus leading to corrosion of the mechanical components, leading to malfunction of the torque limiting function over time.The reason for the partial untight seam welding is the process of welding and subsequent polishing.Icotec has decided to recall all torque wrench devices ref 42-702 as a precautionary measure.
|
| |
|
Event or Problem Description
|
|
A torque wrench was returned from an hospital.During internal testing of the device, it was detected that the torque limiting function of the wrench was not functioning correctly.The torque wrench is used to final tighten the nut screws of the icotec pedicle systems during implantation.
|
| |
|