Catalog Number 74120154
Medical Device Problem Codes Biocompatibility (2886); Migration (4003)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 10/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after thr surgery had been performed on (b)(6) 2008, the patient experienced metallosis and implant mobilization.A revision surgery was performed on (b)(6) 2021 to address this adverse event.Patient's currents status is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Additional information: d10 (modular head added as concomitant device), h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: it was reported that a hip revision surgery was performed due to metallosis and implant mobilization.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers (for the past 12 months) and the reported failure modes to evaluate patterns of repeated failures or defects.No other similar complaints have been identified for the cup and the modular head.However, this will continue to be monitored via routine trending, however it should be noted that these devices are no longer sold.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.As of the date of this medical investigation, the requested supporting clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the metallosis and implant mobilization.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Corrected data: b1 (type of event).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 54MM W/ IMPTR
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18536837
Report Number3005975929-2024-00003
Device Sequence Number6545218
Product Code NXT
UDI-Device Identifier03596010502582
UDI-Public3596010502582
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/02/2012
Device Catalogue Number74120154
Device Lot Number080021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date01/18/2024
Supplement Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2746-2015&Z-0584-20136
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: 74121346 / LOT: 50397.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexUnknown
-
-