| Catalog Number |
450952-NL |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during testing, the anesthesia machine leak check did not pass.A leak was detected.No harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E2 - incorrect due to system limitations.Initial reporter is account sales representative.G5: product code cai is 510k exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 and h6 - evaluation codes: updated.Device evaluation: one device and three photos were received for investigation.A pinhole was observed in the bag in picture 3 and was also seen during visual inspection of the device.The failure mode was confirmed.Based on the analysis conducted in the samples provided, the breathing bags present a damage.Root cause cannot be associated with the manufacturing process since the failure reported could not be reproduced using the tools from assembly process.Root cause cannot be established.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|