| Model Number |
REVEL |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).H10: a third party service technician evaluated the ventilator and found that the tidal volume was inaccurate.As a resolution,the flow sensor was replaced.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the revel ventilator had an inaccurate vt.Furthermore, there was no patient associated with the reported event.
|
| |
|