| Model Number |
N/A |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the battery overheated.This event is related to a malfunction that could potentially lead to a sterility issue/serious injury.However, no patient harm or further outcome was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of the device history record identified no anomalies during manufacturing related to the reported event.Upon receipt, the device does not have power due to overheating.A definitive root cause cannot be determined device is used for treatment.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|