| Model Number |
AU00T0 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
|
| |
|
Event or Problem Description
|
|
A pharmacy student reported failure placement with the intraocular lens (iol) implant device.Additional information was requested, but no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned loose inside the carton.The plunger lock and lens stop have been removed from the device.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted back to mid-nozzle.No damage or abnormalities observed.The lens was returned inside a surgical glove.Viscoelastic was dried on the lens.The lens was removed from the surgical glove and cleaned with klrp for further evaluation.A dimensional inspection was conducted.The lens met specifications per the approved (plan view) template.No lens damage was observed.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.It is unknown if a qualified product was used.The root cause could not be determined for the reported complaint.It is unknown what is meant by "failure placement".The specific issue was not specified.No damage or abnormalities observed with the device.The lens was returned in a surgical glove.The lens was removed from the glove and cleaned with klrp for further evaluation.A dimensional inspection was conducted.The lens met specifications per the approved (plan view) template.No lens damage was observed.Not enough information was provided to determine if a qualified viscoelastic was used.The instruction for use (ifu) instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Follow up attempts were made for more details of the event.No further information has been provided at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|