U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB PS ARTIC SURF SZ6; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB PS ARTIC SURF SZ6; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500506
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Within the procedure, the surgeon removes the trial insert, fb ps tibia trial insert size 6, from the patient who breaks a part of it.The surgery was uneventful, there were no inconveniences or delays and the surgery was accomplished successfully and smoothly.The doctor no longer required the test.There was no impact on the patient and the broken part of the piece remains together with the marked piece inside the instruments.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, within the procedure, the surgeon removes the trial insert, fb ps tibia trial insert size 6, from the patient who breaks a part of it.The surgery was uneventful, there were no inconveniences or delays and the surgery was accomplished successfully and smoothly.The doctor no longer required the test.There was no impact on the patient and the broken part of the piece remains together with the marked piece inside the instruments.The product was not returned to depuy synthes, however photos were provided for review.See attachment source file - reporte de calidad clinica farallones colectivo 452802.The photo investigation revealed that attune fb ps artic surf sz6 is broken from one of the post components.This condition can also be related to the unable to disassemble condition reported as it can't be correctly used if the device is broken.It is also observed that the spring from the broken post is deformed.No other issues were found.This failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface.The attune intuition surgical technique 0612-10-512, emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune fb ps artic surf sz6 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB PS ARTIC SURF SZ6
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18542633
Report Number1818910-2024-01377
Device Sequence Number8914895
Product Code HWT
UDI-Device Identifier10603295132462
UDI-Public10603295132462
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500506
Device Lot NumberMVMHJS130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer01/26/2024
Initial Report FDA Received Date01/19/2024
Supplement Report FDA Received Date01/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-