• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE OY VITAL SIGNS; ANESTHESIA GAS SAMPLING LINE, DISPOSABLE, 3M/10FT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE OY VITAL SIGNS; ANESTHESIA GAS SAMPLING LINE, DISPOSABLE, 3M/10FT Back to Search Results
Model Number 73319-HEL
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: 81 other - at this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It was reported to vyaire medical that the 73319-hel - gas sampling line, 10ft, m/m was partially occluded, causing the anesthesia machine to go into a check defend alarm status.Though there were half-hour delays during the surgery, the customer confirmed that there was no patient harm associated with the reported event.
 
Event or Problem Description
It was reported to vyaire medical that the 73319-hel - gas sampling line, 10 ft, m/m was partially occlkuded, causing the anesthesia machine to go into a check defent alarm status.Though there were half-hour delays during the surgery, the customer confirmed that there was no patient harm associated with the reported event.
 
Additional Manufacturer Narrative
Complaint is not valid and was not confirmed.Investigation of the provided sample showed that the product used was not 73319-hel that the complaint was made against.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SIGNS
Common Device Name
ANESTHESIA GAS SAMPLING LINE, DISPOSABLE, 3M/10FT
Manufacturer (Section D)
VYAIRE OY
kuortaneenkatu 2
la-suomen laani
helsinki uusimaa, fl FIN-0 0510
FI  FIN-00510
Manufacturer (Section G)
VYAIRE OY
kuortaneenkatu 2
qiao long district, tangxia
helsinki uusimaa, fl FIN-0 0510
FI   FIN-00510
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key18544347
Report Number3010838917-2024-00098
Device Sequence Number18106087
Product Code CCL
UDI-Device Identifier10190752117181
UDI-Public(01)10190752117181(10)230167
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K051092
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number73319-HEL
Device Catalogue Number73319-HEL
Device Lot Number230167
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2023
Supplement Date Received by Manufacturer12/26/2023
Initial Report FDA Received Date01/19/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-