Catalog Number CL10041001
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A user facility registered nurse (rn) reported to fresenius that the crit-line clip (clic) device provided inaccurate (low) hematocrit (hct) and hemoglobin (hgb) readings during a patient's hemodialysis (hd) treatment.The clic device was started correctly by staff and the initial readings were reported to be 15.9 hct and 5.4 hgb.Approximately two hours into treatment the readings were retaken and provided as 17.3 hct and 5.9 hgb.The patient's blood was manually drawn and tested and revealed that the patient's hgb was 9.0.The patient did not experience a serious injury nor require medical intervention.The patient's estimated blood loss (ebl) was approximately 5 ml.It was indicated upon initial reporting that the patient was able to complete treatment on the day of the event.Additional information was requested however a response was not received.No parts were returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The complaint investigation did not find objective evidence indicating a product problem, thus the complaint is unconfirmed.
 
Event or Problem Description
A user facility registered nurse (rn) reported to fresenius that the crit-line clip (clic) device provided inaccurate (low) hematocrit (hct) and hemoglobin (hgb) readings during a patient's hemodialysis (hd) treatment.The clic device was started correctly by staff and the initial readings were reported to be 15.9 hct and 5.4 hgb.Approximately two hours into treatment the readings were retaken and provided as 17.3 hct and 5.9 hgb.The patient's blood was manually drawn and tested and revealed that the patient's hgb was 9.0.The patient did not experience a serious injury nor require medical intervention.The patient's estimated blood loss (ebl) was approximately 5 ml.It was indicated upon initial reporting that the patient was able to complete treatment on the day of the event.Additional information was requested however a response was not received.No parts were returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRIT-LINE CLIP (CLIC) - USB
Common Device Name
ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key18546439
Report Number0002937457-2024-00115
Device Sequence Number15723601
Product Code KOC
UDI-Device Identifier00840861101139
UDI-Public00840861101139
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K121599
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCL10041001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by Manufacturer02/22/2024
Initial Report FDA Received Date01/19/2024
Supplement Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-