MAUDE Adverse Event Report: COOK INC; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250); Leak/Splash (1354)
Health Effect - Clinical Code Low Blood Pressure/ Hypotension (1914)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.D1 - brand name: cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set.E3- occupation: materials management admin.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that a catheter from a cook¿s spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray leaked.The device was placed in the patient on (b)(6) 2024.On (b)(6) 2024, leakage at the hub connection was discovered.The complaint device was then removed and replaced.The patient experienced a decreased blood pressure until the device was replaced.No other adverse effects were reported for this incident.
 
Additional Manufacturer Narrative
Correction: b3, h6 (annex a, annex g).It was reported that a catheter from a cook¿s spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray leaked.The device was placed in the patient on (b)(6) 2024.On (b)(6) 2024, leakage at the hub connection was discovered.The complaint device was then removed and replaced.The patient experienced a decreased blood pressure until the device was replaced.No other adverse effects were reported for this incident.Reviews of documentation including the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot and the related subassembly lots revealed one related non-conformance in which one device was scrapped.A complaint history search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The product ifu, [c_t_ctulmabrm_rev7] ¿cook spectrum central venous catheter minocycline/rifampin antibiotic impregnated power injectable,¿ provides the following information to the user related to the reported failure mode: warnings: ¿the safe and effective use or central venous catheters with power injector pressures (safety but-off) set above 325 psi has not been established.Do not power inject if maximum injection rate cannot be verified to meet limit of 10ml/sec.To safely use catheters with a power injector, the technician/heath care professional must verify prior to use that the maximum safety cut-off pressure limit is set at or below 325 psi and that the maximum flow rate is at or below 10 ml/sec.Do not exceed the maximum flow rate of 10 ml/sec.Exceeding the maximum flow rate of 10 ml/sec may result in catheter failure and/or catheter tip displacement.Power injector machine pressure limiting (safety cut-off) settings may not prevent over-pressurization of an occluded catheter.Precautions: ¿catheter should not be used for long-term indwelling applications.¿ based on the information provided, no returned product, and the results of the investigation, cook has concluded that the main cause of failure was a component failure without a design or manufacturing deficiency.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Appropriate measures have been taken to address this failure mode.A capa was previously opened for further investigation.Among other variables, over-engagement of external components onto the hub may contribute to cvc hub cracking.The use of hemostats or lubrication, such as fluids remaining on the hub, during application of external components may further contribute to over-engagement, introducing additional forces to the hub.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18546472
Report Number1820334-2024-00091
Device Sequence Number15699523
Product Code FOZ
UDI-Device Identifier00827002349148
UDI-Public(01)00827002349148(17)251231(10)15496527
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K081113
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberN/A
Device Catalogue NumberC-UTLMY-701J-ABRM-CUSTOM-0042
Device Lot Number15496527
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by Manufacturer05/28/2024
Initial Report FDA Received Date01/19/2024
Supplement Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-