| Model Number |
CCWET0 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, it was noticed that there was defect/crack in lens.The iol was explanted during initial procedure.Additional information has been received stating the iol was exchanged for same model company lens.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned adhered by solution to the non-serialized side of the lens case lid (lens case base not returned).Solution was dried on the lens.The optic was torn, split, cracked, and cut into two portions, typical of insertion and removal.Complaint and product history records were reviewed and documentation indicated the product met release criteria.The file indicated the use of a qualified cartridge and handpiece.The file indicated the use of a viscoelastic that is not qualified for the product combination used.The root cause for the reported complaint appears to be related to failure to follow the ifu.The file indicated the use of a viscoelastic that is not qualified for the product combination used.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|