• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VGXP XP INLK PRI TIB TRAY 75MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VGXP XP INLK PRI TIB TRAY 75MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a knee procedure, the locking bar would not successfully mate with the tibial insert.The procedure was completed using another tibial insert.No adverse events have been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2024-00133 and 0001825034-2024-00134.D10 medical devices: vgxp xp e1 tib brg rl 9x71 catalog#: 195779 lot#: 896670.G2 foreign source: japan.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Up complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned locking identified the locking show signs of attempted implantation and was found to be nicked, gouged, and bent.The tibial tray was not returned as it remains implanted.Device history record was reviewed and no discrepancies were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VGXP XP INLK PRI TIB TRAY 75MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18552388
Report Number0001825034-2024-00134
Device Sequence Number6383251
Product Code MBH
UDI-Device Identifier00880304552807
UDI-Public(01)00880304552807(17)240828(10)456310
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132873
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2024
Device Model NumberN/A
Device Catalogue Number195757
Device Lot Number456310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/25/2023
Supplement Date Received by Manufacturer06/21/2024
Initial Report FDA Received Date01/22/2024
Supplement Report FDA Received Date06/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/28/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-