Model Number 390090
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was explanted due to eri status.No adverse patient events were reported.Should additional information be received, this file will be update.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the device was interrogated.The interrogation could be properly performed and revealed the eri battery status.Next, the icd was subjected to an analysis of the electrical parameters.The current consumption of the icd was checked and found to be normal and as expected.Analysis of the battery condition, however, showed an inconsistency of charge taken from the battery and the battery voltage.A premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.As a first step in the analysis, the device was interrogated.The interrogation could be properly performed and it revealed the eos battery status, detected after explantation of the device.Next, the icd was subjected to an analysis of the electrical parameters.The current consumption of the icd was checked and found to be normal and as expected.Analysis of the battery condition, however, showed an inconsistency of charge taken from the battery and the battery voltage.A premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFORIA 7 VR-T DX DF-1
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18553336
Report Number1028232-2024-00370
Device Sequence Number7597568
Product Code LWS
UDI-Device Identifier04035479127176
UDI-Public(01)04035479127176(17)160229
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2016
Device Model Number390090
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2024
Supplement Date Received by Manufacturer05/03/2024
08/07/2024
Initial Report FDA Received Date01/22/2024
Supplement Report FDA Received Date05/03/2024
08/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age63 YR
Patient SexMale
-
-