|
The device was not received for evaluation; therefore, a device analysis could not be completed; however, the log files associated with the claria device were reviewed.A review of the device logs revealed the programmed min i-drain volume (0ml) was set below the recommended settings in the most recent therapies.The device logs indicated the patient filled with 2544ml during fill 1 of 7 and then subsequently initiated a manual drain of 1536ml.The patient resumed their therapy and drained 2633ml during drain 1.Per the homechoice claria clinician guide, the programmed minimum initial drain should be set to at least 70% of the expected peritoneal volume.If the minimum i-drain volume is set to 0 (zero), it can be associated with a higher risk of iipv due to the possibility of an incomplete drain followed by a full fill volume.Based on the device log analysis, the direct cause of the reported event was determined to be the programmed i-drain alarm being set too low, (use error).Should additional relevant information become available, a supplemental report will be submitted.
|