Model Number 72400025
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 01/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced urinary incontinence.A revision surgery was performed, upon examination fluid leakage due to a hole in the balloon was found.The balloon was explanted and replaced.No additional patient complications were reported.
 
Additional Manufacturer Narrative
There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported patient symptoms are a known risk associated with implant of these devices as indicated in the instructions for use (ifu).Based on the information available, a conclusion code of known inherent risk of device was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 URINARY CONTROL SYSTEM
Common Device Name
DEVICE INCONTINENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18553827
Report Number2124215-2024-02961
Device Sequence Number7597071
Product Code EZY
UDI-Device Identifier00878953000633
UDI-Public00878953000633
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/27/2010
Device Model Number72400025
Device Catalogue Number72400025
Device Lot Number410446017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2024
Initial Report FDA Received Date01/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUFF: UPN: 72400160 / LOT: 420554008; PUMP: UPN: 72400098 / LOT: 419832002
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
Patient SexUnknown
-
-