• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 60MM CUP SIZE REVISION SHELL; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 60MM CUP SIZE REVISION SHELL; PROSTHESIS, HIP Back to Search Results
Catalog Number 00700006020
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: medical product: bone screw self-tapping 6.5 mm dia.30 mm length: catalog#00625006530, lot#j6858515; bone screw self-tapping 6.5 mm dia.30 mm length: catalog#00625006530, lot#j7394928; bone screw self-tapping 6.5 mm dia.35 mm length: catalog#00625006535, lot#j7061001; 0 degree face angle 36mm i.D.Cemented revision shell liner use with 60mm o.D.Shell: catalog#00710506036, lot#64955355.G2: foreign: germany.Multiple mdr reports have been filed for this event.Please see associated reports: 0001822565-2024-00220; 0001822565-2024-00221; 0001822565-2024-00222.Diligence is in process to determine whether the product is available for evaluation.The investigation is in progress.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a total hip procedure, the screws used to attach the cup did not seat and drilled completely through.There was a surgical delay of two (2) hours while the screws and shell were removed and replaced with a second set of implants.No patient impact has been reported as a result of the malfunction.Due diligence is in progress for this event; to date no further information has been reported.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned devices identified three of seven screw holes of the shell are damaged.One of the damaged holes appears to have been deformed in a manner that has increased its¿ diametrical appearance.Discoloration is seen on the rim feature.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: an initial tha was performed on an unknown date with unknown product.The patient was revised due to cup loosening with acetabular bone defects and cysts.When trying to place the tm shell, the screw went through the shell.A different shell was used to complete the procedure.The root cause of the reported issue is attributed to user error, as the surgeon used incompatible screws with the tmars shell, causing them to pull through and damage the shell.Per the ifu, it states not to use this product for other than labeled indications (off label use).A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
60MM CUP SIZE REVISION SHELL
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18554448
Report Number0001822565-2024-00218
Device Sequence Number6402395
Product Code JDI
UDI-Device Identifier00889024310896
UDI-Public(01)00889024310896(17)280217(10)65641020
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K001759
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00700006020
Device Lot Number65641020
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by Manufacturer03/25/2024
Initial Report FDA Received Date01/22/2024
Supplement Report FDA Received Date04/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Patient SexPrefer Not To Disclose
-
-