Model Number TV-AB2680-N
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 12/29/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.H3 other text : device remains implanted.
 
Event or Problem Description
The patient was being treated for an abdominal aortic aneurysm with the alto stent graft system.The patient had a short off label neck with a severe angle, making them unsuitable for a z-fen procedure.Despite challenges in imaging and device deployment due to the neck's angle, the procedure was successful in sealing the aneurysm.A post-procedure issue arose as the left renal was lost, possibly due to the migration of the ring or inaccuracies in marking.The patient is stable, and the case outcome, while successful in sealing the aneurysm, resulted in the loss of the left kidney.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good-faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the alto, displaced implant with left renal occlusion complaint is confirmed.This is consistent with the reported adverse event/incident.It was reported that the neck was only 5mm long and it was highly angulated.This could have contributed to the reported event, but that could not be conclusively determined.The device, user, procedure, or anatomy-relatedness of this complaint could not be determined.The procedure-related harms were left renal infarction and conversion to brachial access.The final patient status was reported as discharged to home on postoperative day three in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTO
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
,
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
,
santa rosa, CA 95403
8009832284
MDR Report Key18557053
Report Number3008011247-2024-00008
Device Sequence Number12694791
Product Code MIH
UDI-Device Identifier00850007370893
UDI-Public(01)00850007370893(17)26091110FS090723-23
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/11/2026
Device Model NumberTV-AB2680-N
Device Lot NumberFS090723-23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/29/2023
Supplement Date Received by Manufacturer02/27/2024
Initial Report FDA Received Date01/22/2024
Supplement Report FDA Received Date03/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION IX ILIAC LIMB FS042023-25; OVATION IX ILIAC LIMB FS071723-05; OVATION PRIME FILL POLYMER FF052622-02
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexFemale
-
-