| Catalog Number |
2C6204 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a customer was unable to draw blood through a large bore stopcock.This occurred during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: this event occurred during patient use.An alternative means of drawing blood was used to resolve the event.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the no flow.A pressure test was performed on the sample, and the sample was found to be within the product specification.The reported condition was not verified.According to the instruction for use (ifu) manual, the device is used for fluid administration, not fluid draw.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|