• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number RONYX35030UX
Medical Device Problem Codes Material Deformation (2976); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
An attempt was made to use one resolute onyx coronary drug eluting stent to treat a severely tortuous, severely calcified lesion with 100% chronic total occlusion (cto) in the distal right coronary artery (rca). the device was not inspected.Negative prep was not performed.The device did not pass through a previously deployed stent.Resistance was noted while advancing the device to the lesion.Excessive force was not used.It was reported that the device failed to cross the lesion. the lesion was treated by using a cto wire and balloon to re-prepare the lesion.Then a same size resolute integrity stent was deployed with no issues noted.The case was successful, and the patient is good.The patient is alive with no injury.
 
Additional Manufacturer Narrative
Product analysis: the device was returned to medtronic for analysis.The stent was positioned on the balloon between the marker bands as per position specification, but due to proximal and distal stent deformation the stent did not meet visual acceptance specification.Deformation was evident to the proximal and distal stent wrap with struts raised, stretched and overlapping.Deformation evident to the distal tip.The inner lumen patency was verified with a mandrel.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18561996
Report Number9612164-2024-00360
Device Sequence Number6660784
Product Code NIQ
UDI-Device Identifier00643169557116
UDI-Public00643169557116
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2025
Device Model NumberRONYX35030UX
Device Catalogue NumberRONYX35030UX
Device Lot Number0011163021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2024
Initial Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
Patient Weight62 KG
-
-