Catalog Number CDC-25142-XCN1A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
We elected to place a temporary hd catheter in the femoral vein on the r because of access issues.We were able to easily access the vein using us guidance.We thread the wire and confirmed venous placement.A large skin incision was made to accommodate the dilators and catheter.The problem began upon dilation.We encountered more resistance than expected during dilation and upon trying to advance the catheter on the first attempt.I attempted to advance the catheter myself but was unsuccessful.At that time, i elected to pull the catheter out, change try another kit.I threaded a catheter over the wire, removed the old wire which had a very small kink and replace it with a new wire.I then re-dilated with both dilators and then tried to advance the new 20cm 14f hd catheter.I was unsuccessful.I met a significant amount of resistance, approximately 8-10cm in, and despite many maneuvers to advance, i was unsuccessful.The catheter was too soft and would start to kink under the skin.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.
 
Additional Manufacturer Narrative
(b)(4).The customer report of dilator tip damage was confirmed by visual inspection of the customer supplied photos.A full complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "do not use excessive force when introducing guidewire or tissue dilator as this can lead to vessel perforation, bl eeding, or component damage." a device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
We elected to place a temporary hd catheter in the femoral vein on the r because of access issues.We were able to easily access: the vein using us guidance.We thread the wire and confirmed venous placement.A large skin incision was made to accommodate the dilators and catheter.The problem began upon dilation.We encountered more resistance than expected during dilation and upon trying to advance the catheter on the first attempt.I attempted to advance the catheter myself but was unsuccessful.At that time, i elected to pull the catheter out, change try another kit.I threaded a catheter over the wire, removed the old wire which had a very small kink and replace it with a new wire.I then re-dilated with both dilators and then tried to advance the new 20cm 14f hd catheter.I was unsuccessful.I met a significant amount of resistance, approximately 8-10cm in, and despite many maneuvers to advance, i was unsuccessful.The catheter was too soft and would start to kink under the skin.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AHDC AGB KIT: 2-L 14 FR X 20 CM
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18562015
Report Number9680794-2024-00018
Device Sequence Number8123795
Product Code NIE
UDI-Device Identifier10801902213553
UDI-Public10801902213553
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K143102
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/13/2025
Device Catalogue NumberCDC-25142-XCN1A
Device Lot Number33F23G0115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2023
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexMale
-
-