• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC DOMINICANA CSF-VENTRICULAR PORT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC DOMINICANA CSF-VENTRICULAR PORT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR Back to Search Results
Model Number 44010
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a ventricular port.It was reported that when the doctor was performing chemotherapy port placement on the patient, the package was opened, and it was found there was a crack at the connection between the reservoir and the ventricular end tube.The device was replaced for a new one to complete the surgery.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
H3.Product analysis determined that the visual and functional inspection confirmed the port had been damaged/cracked and leaked when tested.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-VENTRICULAR PORT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR
Manufacturer (Section D)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR  11500
Manufacturer (Section G)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR   11500
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key18562504
Report Number9612501-2024-00202
Device Sequence Number6660799
Product Code LKG
UDI-Device Identifier00763000030292
UDI-Public00763000030292
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K874468
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number44010
Device Catalogue Number44010
Device Lot Number0225622325
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2024
Supplement Date Received by Manufacturer03/28/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date03/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
-
-