| Catalog Number |
254500720 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
When the attune femoral specimen is broken, the condyle has broken off.Was surgery delayed due to the reported event? no, was procedure successfully completed? yes, were fragments generated? yes, if yes, were they removed easily without additional intervention? yes.
|
| |
|
Event or Problem Description
|
|
Additional information was received: a.Were all pieces retrieved from the patient? no piece entered the patient's body b.Were there any adverse consequences that affected the patient because of the reported event? no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)94).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : when the attune femoral specimen is broken, the condyle has broken off.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of the returned sample revealed that the attune cr fem trial sz 10 lt was found broken from the right condyle.The broken fragment was returned for examination.No other issues were observed.A dimensional inspection was unable to be performed due to post manufacturing damage.The fracture surface is consistent with overload due to a combination of heavy impaction and the trial fitting too tight over the posterior/anterior aspect of the resected femur bone forcing the curved features of the femoral trial to open.The combination of heavy impaction and possible discrepancies in the resection depth between the femur and the trial suggest unintended user error.Furthermore, per the attune surgical technique (dsus/jrc/0316/1437 rev.K) it is cautioned when extracting the trial, rocking the trial medio-laterally may cause condylar fracture.Such rocking should be avoided.The overall complaint was confirmed as the observed condition of the attune cr fem trial sz 10 lt would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|