| Catalog Number |
MX960 |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pressure sensor did not respond after being connecting to the sensor board and the pressure value could not be detected.After replacing the pressure sensor board with a new one, the display appeared.This reportedly occurred with 43 pressure sensors.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review, this was found to be a non-reportable event.The event does not meet the definition of a reportable death or serious injury as the malfunction of the device did not cause or contribute to a patient death, a life-threatening event, permanent impairment of a body function or structure, required medical or surgical intervention to preclude serious impairment of a body structure or function.Additionally, the reportable malfunctions list indicates that the device in the reported incident would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Therefore, mfr#1526863-2024-00003-00 will be cancelled.
|
| |
|