|
There was an allegation of questionable gen.2 ise indirect for na results for approximately 30 patient samples on a cobas 8000 cobas ise module.The customer provided an example of one patient sample with discrepant na results.The customer noticed moving result averages were drifting low.The customer re-calibrated, replaced the na electrode, replaced ise reagents, and repeated controls, but there was no change.The customer then repeated the affected patient samples on a second ise module the initial na result was 136 mmol/l.The sample was repeated on another ise module and the result was 143 mmol/l.The repeat result was deemed correct.
|