EXACTECH, INC. NV GXL LINR, NTRL, 36MM ID, GROUP 4 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
|
Back to Search Results |
|
| Catalog Number |
130-36-54 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
01/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient had a right total hip arthroplasty on (b)(6) 2014 and then experienced a surgical revision approximately 9 years, 2 months later on (b)(6) 2024.The patient had a recalled hip liner.Reported event is not related to breakage of a device.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.D10.Concomitants: (b)(6), 170-36-00 - biolox delta femoral head 36mm od, +0mm; (b)(6), 180-65-30 - alteon 6.5mm screw, 30mm; (b)(6), 186-01-60 - integrip cc, cluster 60mm, g4; (b)(6), 188-01-10 - wedge plasma x/o sz 10.These devices are used for treatment not diagnosis.Pending investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.Updated/additional information ¿ g1.G2.G4.H6.The most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this cannot be not confirmed with the information provided; devices were not returned.There is no other information available.These devices are used for treatment not diagnosis.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3/g4, h6.The following sections were corrected: h6.Mdr section codes updated/corrected: b, c, e, g.The revision reported may have been due to a combination of risk factors: such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential) including but not limited to use error, implant positioning, and patient factors.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|