SYNTHES GMBH 11.5MM/8.5MM PROTECTION SLEEVE FOR RECON LOCKING; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.010.075
Medical Device Problem Codes Device-Device Incompatibility (2919); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2024 during a procedure, the protection sleeve for the head elements of the femoral recon nailing system did not line up with the holes in the aiming arm.The surgeon needed to mallet the protection sleeves in place and this caused damage to the aiming arm.There were no adverse events and the case was completed without incident.The procedure was successfully completed with no delay.No patient involvement was reported.This report is for a 11.5mm/8.5mm protection sleeve for recon locking.This is report 3 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.5MM/8.5MM PROTECTION SLEEVE FOR RECON LOCKING
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18564602
Report Number8030965-2024-01219
Device Sequence Number15483083
Product Code FZX
UDI-Device Identifier10886982067456
UDI-Public(01)10886982067456
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer01/25/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date02/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
11.5/8.5 PROTEC SLEEVE FOR RECON LCKNG; RAD AIMING ARM/FRN GREATER TROCHANTER
Patient SexUnknown
-
-