• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON JET PLUS TAP-ON (DNA) G137; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON JET PLUS TAP-ON (DNA) G137; SCALER, ULTRASONIC Back to Search Results
Catalog Number G137
Medical Device Problem Codes Restricted Flow rate (1248); Overheating of Device (1437); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
While using a cavitron jet plus g137, they allege that they have minimal water and the handpiece is heating up, no injury resulted.
 
Additional Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803 the device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Additional Manufacturer Narrative
Root cause: emh hp;cable,conn/gun cable damaged.Debris buildup in the handpiece cable and leaking water around the water turn knob continuous water leaking due to debris buildup in the water system not allowing the solenoid to close completely.Jet-mate has a bent contact pin.Manifold has bad regulation resulting in inconsistent powder flow.Cracked powder bowl (has a different screw that is going through the top cover of powder bowel).Baffle is damaged.No operation with the wireless foot pedal.No powder flow due to a clogged y-fitting.Debris build in the water filter.Note: will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.(this unit looks like it has been previously open and worked on).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON JET PLUS TAP-ON (DNA) G137
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key18564643
Report Number2424472-2024-00003
Device Sequence Number17856978
Product Code ELC
UDI-Device IdentifierD00381614281
UDI-PublicD00381614281
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K052334
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG137
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer01/23/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date02/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1