| Catalog Number |
03P86-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Cardiac Enzyme Elevation (1838)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, abbott point of care was contacted by a customer regarding act celite cartridges that yielded descrepant results on a 52 year old male patient presented in the er as stemi.There was no additional patient information available at the time of this report.Return product is available for investigation.I-stat date tested result heparin time comment 366036 on (b)(6) 2023 03:51 201 sec <
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: #(b)(4).The investigation was completed on 14-mar-2024.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Am (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|